Electrical connection Liquistation CSF48
44 Endress+Hauser
LEDs on front of module
LED Designation Color Description
PWR Power GN Supply voltage is applied and module is initialized.
BF Bus failure RD Bus failure
SF System failure RD Device error
COM Communication YE PROFIBUS message sent or received.
T Bus termination YE • Off = No termination
• On = Termination is used
DIP switches on front of module
DIP Factory setting Assignment
1-128 ON Bus address (→ "Commissioning/communication")
OFF Write protection: "ON" = configuration not possible via the bus, only via local
operation
Service OFF The switch has no function