Liquistation CSF48 System integration
Endress+Hauser 65
Operation
Clicking a menu name or a function corresponds to pressing the navigator.
You can make your settings conveniently via the computer keyboard.
Instead of using an Internet browser, you can also use FieldCare for configuration via
Ethernet. The Ethernet DTM required for this is an integral part of the "Endress+Hauser
Interface Device DTM Library".
Heartbeat verification
You can also start Heartbeat verification via the web server. This has the advantage that you
can view the results directly in the browser and can avoid using an SD card.
1. Open the menu: Diagnostics/System test/Heartbeat.
2. Perform verification.
3. Either Verification results (quick display and export to SD card) or Additional
Functions (additional menu below the limit line ).
4. Additional Functions/Heartbeat: Select the language of the pdf file.
The verification report is displayed in the browser and can be printed out, saved as a
pdf file, etc.
8.1.2 Service interface
You can connect the device to a computer via the service interface and configure it using
"FieldCare" . Furthermore, configurations can also be saved, transferred and documented.
Connection
1. Connect the service connector to the interface on the Liquiline base module and connect
it to the Commubox.
2. Connect the Commubox via the USB connection to the computer on which FieldCare is
installed.
FieldCare
Service
FXA291
A0039618
 66 Connection overview
Establishing the data connection
1. Start FieldCare.