EasyManua.ls Logo

Endress+Hauser Liquistation CSF48 - Page 100

Endress+Hauser Liquistation CSF48
268 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation Liquistation CSF48
100 Endress+Hauser
Path: Menu/Setup/Sampling programs
Function Options Info
Current program: Read only The last sampling program to be created or used is
displayed.
Status Read only Display "Active":
The sampling program has been started and the
device takes a sample as per the set parameters.
Display "Inactive":
No sampling program has been started, or a
program that was running has been paused.
Display "Pause":
Sampling program paused.
Setup program
New A list of all the programs created is displayed. For
this reason, it is often helpful to add a "B" for Basic in
the program name.
Program1, which is supplied with the device, is displayed, as is a list of all the programs already created (Basic, Standard
or Advanced programs). You can either create a new program or select an existing one.
If you select an existing program, you can edit, delete, start or duplicate it. Furthermore, you can also see whether this
program is a Basic, Standard or Advanced program.
If you are creating a new program, select the Basic, Standard or Advanced program type.
Basic
Program name Customized text Use a distinct name for your sampling program. The
program name can be up to 16 characters long.
Bottle configuration Choice of all possible bottle
combinations
The ordered bottle configuration is preset or the
configuration selected in the setup is displayed.
Options:
- 1x - PE direct distribution
- 2x - PE direct distribution
- 4x - PE direct distribution
1)
- 4x - glass direct distribution
- 12x - PE direct distribution
- 12x - PE/glass distributor plate
- 24x - PE direct distribution
- 24x - PE/glass distributor plate
- 6x+1x - PE/glass distributor
plate
- 6x+2x - PE+PE distributor
plate
- 6x+2x - PE+glass distributor
plate
- 12x+1x - PE/glass distributor
plate
- 12x+2x - PE+PE distributor
plate
- 12x+2x - PE+glass distributor
plate
- 12x+6x PE direct distribution
- 12x+6x PE/glass distributor
plate
1)
Use this option for bottle configuration "W" (4 x
5000 ml Schott bottles, DURAN GLS 80).

Table of Contents

Other manuals for Endress+Hauser Liquistation CSF48

Related product manuals