Operation Liquistation CSF48
140 Endress+Hauser
Programming example: Event program
The following example explains how to program an event program with time-paced sampling
into a composite container and time-paced sampling into individual bottles. The composite
container is emptied daily.
a0014126s
Fig. 89: Event-controlled sampling
ApH value: 8.0
BpH value
1 Sampling into individual bottles when pH value > 8.0
2 Sampling into composite container when pH value < 8.0
A
B
1
1
2
8.0
Path: Menu/Setup/Sampling programs/Setup program/New
Function Entry for sample program
Advanced
Program name TDTE 12+20
Bottle configuration 12x + 1x
Bottle volume left 1000 ml
Bottle volume right 2000 ml
Start condition Immediate
Stop condition Continuous
Bottle assignment Statical bottle assignment
Setup subprogram: New
Programpart TD1
Sampling mode Time paced CTCV
Sampling interval 15 min
Sampling volume 100 ml