Liquistation CSF48 Operation
Endress+Hauser 141
Multiple bottles 1
Enable subprogram Immediate
Sample at enable Yes
Synchronize samplings To subprogram start
-> "SAVE" -> "ESC"
Setup subprogram: New
Programpart EE1
Sampling mode Time paced CTCV
Sampling interval 6 min
Sampling volume 100 ml
Bottle change mode
Only if Bottle assignment = "Statical bottle
assignment" is selected in the main program
Number of samples
Samples per bottle 10
Multiple bottles 0
Enable subprogram Event
Start condition No delay
Activation event
Number of events 1
Event Editor 1
Source of data CH1: 1:1 pH glass
Measured value pH
Operating mode Upper limit
Limit value pH 8.00
Hysteresis pH 0.20
Start delay 0 s
Switch off delay 0 s
Sample at enable Yes
Sample at disable No
Deactivation Enable invalid
Synchronize samplings To subprogram start
Path: Menu/Setup/Sampling programs/Setup program/New
Function Entry for sample program