Press the "SAVE" soft key to save the program. In the overview, you can start the program created
under "Select sampling program".
Programming example: Limit value-controlled program
The following example explains how to program a limit value-controlled sampling program with
flow-paced sampling into individual bottles above the limit value and time-paced sampling into
individual bottles below the limit value. When the subprogram changes, the system switches to
the next bottle.
-> "SAVE" -> "ESC" -> "ESC"
Bottle assignment Statical bottle assignment
Bottle assignment table
Bottle 1
...
Bottle 12
Bottle 13
EE1
EE1
TD1
-> "ESC"
Path: Menu/Setup/Sampling programs/Setup program/New
Function Entry for sample program
Advanced
Program name UVT 12x3
Bottle configuration 12x
Bottle volume 3000 ml
Start condition Date/time
Start date DD.MM.YYYY e.g. 03.06.2010
Start time HH:MM:SS e.g. 12:00:00
Stop condition Continuous
Setup subprogram: New
Programpart VD1
"SAVE" - saves the "VD1" subprogram
Setup subprogram: New
Programpart ZD1
"SAVE" - saves the "ZD1" subprogram
Path: Menu/Setup/Sampling programs/Setup program/New
Function Entry for sample program