Liquistation CSF48 Operation
Endress+Hauser 143
"ESC" - the main program is displayed
Bottle assignment Dynamical bottle assignment
Setup subprogram: VD1
Edit
Sampling mode Flow paced VTCV
Flowmeter input Current input S:1 (max. 15 samples/h = 4 min)
Sampling interval 50 m³
Sampling volume 100 ml
Bottle change mode Time
Time interval 06:00
Multiple bottles 0
Enable subprogram External activation
Start condition No delay
Activation event
Number of events 1
Event Editor 1
Source of data Current input S:1
Measured value Flow rate
Operating mode Upper limit
Limit value 155 l/s
Hysteresis 5 l/s
Start delay 0 s
Switch off delay 0 s
"ESC" - "ESC"
Sample at enable Yes
Sample at disable No
Stop condition Enable invalid
Synchronize samplings To subprogram start
"SAVE" - "ESC"
Setup subprogram: ZD1
Edit
Path: Menu/Setup/Sampling programs/Setup program/New
Function Entry for sample program