Liquistation CSF48 Operation
Endress+Hauser 179
10.6.4 Measuring range switch
A measuring range switching (MRS) configuration includes the following options for each of the
four binary input states:
Mode of operation (conductivity or concentration)
Concentration table
Temperature compensation
Current output turndown
Limit contactor range
If an MRS set is assigned to a channel and MRS is enabled for this set, the measuring range
configuration selected via the binary inputs is used instead of the normal configuration of the
linked sensor channel. For the current outputs and limit contactors to be controlled by the MRS,
they must be connected to the MRS set and not to the measuring channel.
Current outputs and limit contactors that are connected to an MRS set receive the measured
value corresponding to the mode of operation, as well as the corresponding turndown or range
for limit value monitoring, via this MRS set. A limit contactor connected to an MRS set always
uses the "Out of range check" mode, i.e. it switches if the value is outside the configured range.
If a current output or limit contactor is connected to an MRS set, the turndown, monitoring
range and limit contactor mode can no longer be configured manually. Therefore, these options
are hidden in the current output menu and limit contactor menu.
Example: CIP cleaning in a brewery
Beer Water Base Acid
Binary input 1 0011
Binary input 1 0101
Measuring range
00
Measuring range
01
Measuring range
10
Measuring range
11
Operating mode Conductivity Conductivity Concentration Concentration
Conc. Table - - NaOH 0..15% User table 1
Compensation User table 1 Linear - -
Current output
Range lower value 1.00 mS/cm 0.1 mS/cm 0.50 % 0.50 %
Range upper value 3.00 mS/cm 0.8 mS/cm 5.00 % 1.50 %
Limit switches
Range lower value 2.3 mS/cm 0.5 mS/cm 2.00 % 1.30 %
Range upper value 2.5 mS/cm 0.7 mS/cm 2.10 % 1.40 %