EasyManua.ls Logo

Endress+Hauser Liquistation CSF48 - Page 18

Endress+Hauser Liquistation CSF48
268 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Installation Liquistation CSF48
18 Endress+Hauser
5.1.2 Mounting location
For version with pump
1. Correct
The suction line must be routed with a downward slope to the sampling point.
2. Incorrect
The sampler should never be mounted in a place where it is exposed to aggressive gases,
e.g. hydrogen sulfide (H
2
S).
3. Incorrect
Avoid siphoning effects in the suction line
4. Incorrect
The suction line should never be routed with an upward slope to the sampling point.
Note the following when mounting the sampler:
1. The ventilation of the mounting location must be guaranteed.
2. Place the device on a level surface.
3. Protect the sampler from additional heating (e.g. from heaters).
4. Protect the sampler from mechanical vibrations.
5. Protect the sampler from strong magnetic fields.
6. Make sure air can circulate freely through the side panels of the housing. Do not mount the
sampler directly against a wall. Allow at least 150 mm (5.9") between the wall and the
sampler on the left and right-hand sides.
a0011693
Fig. 10: Mounting conditions for Liquistation CSF48 for open channels

Table of Contents

Other manuals for Endress+Hauser Liquistation CSF48

Related product manuals