EasyManua.ls Logo

Endress+Hauser Liquistation CSF48 - Page 181

Endress+Hauser Liquistation CSF48
268 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Liquistation CSF48 Operation
Endress+Hauser 181
Conc. Table Options
NaOH 0..15%
HCl 0..20%
HNO3 0..25%
HNO3 24..30%
H2SO4 0..28%
H2SO4 40..80%
H2SO4 93..100%
H3PO4 0..40%
NaCl 0..26%
•User table 1
•User table 2
•User table 3
•User table 4
Factory setting
NaOH 0..15%
Concentration tables saved at the factory:
NaOH: 0 to 15%, 0 to 100 ˚C (32 to 212 ˚F)
HCl: 0 to 20%, 0 to 65 ˚C (32 to 149 ˚F)
HNO
3
: 0 to 25%, 2 to 80 ˚C (36 to 176 ˚F)
H
2
SO
4
: 0 to 28%, 0 to 100 ˚C (32 to 212 ˚F)
H
2
SO
4
: 40 to 80%, 0 to 100 ˚C (32 to 212 ˚F)
H
2
SO
4
: 93 to 100%, 0 to 100 ˚C (32 to 212 ˚F)
H
3
PO
4
: 0 to 40%, 2 to 80 ˚C (36 to 176 ˚F)
NaCl: 0 to 26%, 2 to 80 ˚C (36 to 176 ˚F)
Compensation Options
•None
•Linear
NaCl (IEC 746-3)
Water ISO7888 (20°C)
Water ISO7888 (25°C)
•UPW NaCl
•UPW HCl
•User table 1
•User table 2
•User table 3
•User table 4
Factory setting
Linear
Various methods are available to compensate for the
temperature dependency.
Depending on your process, decide which type of
compensation you want to use.
Alternatively, you can also select "None" and thus
measure uncompensated conductivity.
Current output
Range lower value 0 to 100 % Operating mode = "Concentration"
Range upper value
Range lower unit Options
•S/m
μS/cm
•mS/cm
•S/cm
μS/m
•mS/m
Factory setting
mS/cm
Operating mode = "Conductivity"
Define the units and range limits for the output values.
Range lower value Enter value
Range upper unit Enter value
Range upper value Enter value
Path: Menu/Setup/Additional functions/Measuring range switch
Function Options Info

Table of Contents

Other manuals for Endress+Hauser Liquistation CSF48

Related product manuals