Diagnostics and troubleshooting Liquistation CSF48
192 Endress+Hauser
339 Liquidsensor M On Off Sensor LF1 fouled
-> Clean the sensor soon
-> The sensitivity can be configured under
Setup/General settings/Sampling/Conductive
sensor
340 Liquidsensor M On Off Sensor LF1 fouled
-> Clean sensor
-> The sensitivity can be configured under
Setup/General settings/Sampling/Conductive
sensor
343 Power supply M On Off Power supply failure
344 Program pause C On Off Sampling program paused
345 Time changeover M On Off Daylight saving time/winter time setting
Normal time (winter time) active.
346 Time changeover M On Off Daylight saving time/winter time setting
Daylight saving time active
347 Sample confirm. F On On Sampling command has not been processed
-> Check internal cable to 1IF
-> Reset the software
348 Read program F On On Selected program cannot be read from the
program memory
--> Create new program
349 Write program F On On Program created cannot be saved
Hardware error has occurred
--> Contact the Service Department
351 Delete prog. F On On Selected program cannot be deleted from the
program memory
--> Reset the software
352 Read Prog.list F On On Program list cannot be read from the program
memory
--> Perform device reset under
Diagnostics/System test/Reset/Device reset
353 Overfill check F On Off Total capacity of bottle reached
• No further sampling to current bottle is
triggered
-> Adjust the sampling program under Program
selection if desired
354 Bottle check F On Off No empty bottles available for current program
• No further sampling
-> Check the program settings under Program
selection
355 Start time over M On Off Start time entered is in the past
• Enter a new start time
No. Message Factory settings Tests or remedial measures
Cat. Diag.
on/off
Error
current