Liquistation CSF48 Diagnostics and troubleshooting
Endress+Hauser 193
356 Overfill check F On Off The total sample volume does not fit in the
sample bottle
-> Change the sample volume
357 Sampling faulted M On Off Sample discarded
There are too many sampling requests pending
-> Adjust the sampling program under Program
selection
358 Configuration F On On Program configuration does not match the
current device configuration
-> Adjust the configuration
359 Emptying error F On On Error when draining
Draining and sampling program canceled
-> Check connection to FMSY1 module
-> Check 4R module, replace if necessary
-> Restart the software under
Setup/Diagnostics/System test/Restart
366 Module connect. F On On No communication with actuator module
-> Check internal connecting cable to module 1IF
370 Intern. Voltage F On On Internal voltage outside the valid range
--> Check supply voltage
373 Electr. temp. M On Off High electronics temperature
--> Check ambient temperature and energy
consumption
374 Sensor check F On Off No measurement signal from sensor
-> Check sensor connection
-> Check sensor, replace if necessary
375 No 4R module F On On No connection to 4R module
-> Check 4R module, replace if necessary
-> Restart the software under
Setup/Diagnostics/System test/Restart
401 Reset to default F On On Factory reset is performed
406 Param. active C Off Off --> Wait for configuration to be finished
407 Diag. active C Off Off --> Wait for maintenance to be finished
412 Writing backup F On Off --> Wait for the write process to be finished
413 Reading backup F On Off --> Wait
No. Message Factory settings Tests or remedial measures
Cat. Diag.
on/off
Error
current