Liquistation CSF48 Diagnostics and troubleshooting
Endress+Hauser 211
Overcurrent Read only No: no error
Yes: the heating is defective
-> Contact the Service Department
Overcurrent Read only No: no error
Yes: the fan is defective
-> Contact the Service Department
Sample compartment Read only The current temperature of the sample compartment is
displayed.
Sample compartment Read only When you start the heating test, the temperature at the
start time is displayed
Heating test offor Heating test on -> progress of test is displayed
Start Test Action Start the heating test.
Stop test Action End the heating test.
Manual sampling
Bottle configuration Read only
Bottle volume Read only
Distributor position Options
•Front
• Bottle 1
...
•Back
Select which bottle should be filled with the sample.
Sample volume 50 to 2000 ml
Factory setting
100 ml
You can change the sample volume in the version with
the peristaltic pump.
Sample volume Factory setting
200 ml
The sample volume is preset at the factory in the version
with the vacuum pump.
Start sampling Action
Peristaltic pump(only for version with peristaltic pump)
Pump purge Action
Pump purge, to stop
press ESC
Read only
Current pump run time Read only
Power supply Read only The current supply voltage is displayed.
With AC power supply: 24 V ±0.5 V
With DC power supply: 22 to 28 V
Motor current Read only The current consumption of the pump is displayed.
Diagnostics/Systemtest/Reset
Function Options Info