Liquistation CSF48 Operation
Endress+Hauser 81
Example for setting up a new data logbook (for sensors)
1. Menu/Setup/General settings/Logbooks/Data logbooks/New:
a. Logbook name: Assign a name, e.g. "01".
b. Source of data: Select a data source, e.g. the sensor connected to binary input 1.
c. Measured value: Select the measured value that you want to record.
d. Scan time: Specify the interval between two logbook entries.
e. Data logbook: Activate the logbook. Specify the type of memory, "Ring buffer" or "Fill up
buffer".
Data logbook Options
•Off
Ring buffer
Fill up buffer
Factory setting
Off
Ring buffer
If the memory is full, the latest entry automatically
overwrites the oldest entry.
Fill up buffer
If the memory is 80% full, the device displays a
diagnostic message.
If the memory is full, there is an overflow, i.e. no new
values can be saved. The controller displays a
corresponding diagnostic message. The memory then
has to be cleared manually.
Line plotter Menu to define the graphic display
Axes Options
•Off
•On
Factory setting
On
Should the axes (x, y) be displayed (On) or not (Off)?
Orientation Options
Horizontal
•Vertical
Factory setting
Horizontal
You can choose whether the value curves should be
displayed from left to right ("Horizontal") or from top to
bottom ("Vertical").
If you want to display two data logbooks simultaneously,
make sure that both logbooks have the same settings
here.
X-Description Options
•Off
•On
Factory setting
On
Decide whether a description should be displayed for the
axes and whether grids should be shown. In addition,
you can also decide whether a pitch should be displayed.
Y-Description
Grids
Pitches
X Pitch/Grid distance 10 to 50%
Factory setting
10 %
Specify the pitch.
Y Pitch/Grid distance
Remove Action This action removes the data logbook. Any data that
have not been saved are lost.
Path: Menu/Setup/General settings/Logbooks
Function Options Info