Operation Liquistation CSF48
82 Endress+Hauser
2. ../Finished: Execute this action.
--> Your new logbook now appears in the list of data logbooks.
3. Select the data logbook with the name "01".
4. If you selected "Fill up buffer", you can also decide whether you want to receive a diagnostic
message in the event of memory overrun.
5. Depending on the type of memory selected, you receive information about the memory
space (for "Ring buffer") or the time remaining until memory overrun (for "Fill up buffer").
6. Define the graphic display mode in the "Line plotter" submenu.
10.2.5 Configuring the sampling depending on the device version
The list of functions displayed depends on the device version selected.
For version with vacuum pump:
1)
For version with peristaltic pump:
2)
For version with distribution arm drive:
3)
For version with sampling assembly:
4)
Path: Menu/Setup/General settings
Function Options Info
Sampling
Number of bottles Choice of all possible bottle
combinations
The bottle configuration you ordered is preset in
the device.
Bottle volume 0 to 100000 ml
Factory setting
Depends on the bottle
configuration
Distribution parking
3)
Options
•Back
•None
Factory setting
Back
Causes the distribution arm to go to the center at
the back or remain parked in the current position
when the device is started or the program is ended.
Distribution reference
3)
Options
Pre sampling
Pre bottle change
Pre program start
Factory setting
Pre sampling
The distribution arm goes through a reference point
depending on the option selected.
Pre sampling:
This means that the distribution arm performs a
reference test before each individual sampling.
Pre bottle change:
This means that the distribution arm performs a
reference test in every subprogram.
Pre program start:
This means that a single reference test is performed
before the program starts.