Liquistation CSF48 Operation
Endress+Hauser 83
Power failure Options
Resume program
•Stop program
Factory setting
Resume program
Decide how the sampler should react when it is
energized after a power failure.
Resume program:
Time and flow-paced
The program calculates the omitted samples and
enters them in the logbook as failed. When the
program is restarted, it continues where it was
interrupted.
•Flow-paced
No samples are entered in the logbook during
the power failure. When the program is
restarted, it continues where it was interrupted.
Sample retries
1), 2), 3)
0 to 3
Factory setting
0
If sampling is started and no sample is drawn in,
sampling can be repeated up to 3 times.
Sampling delay 0 to 99 s
Factory setting
0 s
The start of the sampling cycle can be delayed by up
to 99 s. The binary output is switched without any
delay.
Liquid detection
1)
Options
•Automatic
Semi automatic
Factory setting
Automatic
If "Semiautomatic" is selected, the purge times and
intake times can be defined separately.
Dosing volume
1), 4) 1)
20 to 350 ml
Factory setting
200 ml
4)
10 to 1000 ml
Factory setting
200 ml
1)
Adjust the dosing tube in the dosing chamber to
change the dosing volume. The level in the bottle is
calculated using the set dosing volume.
4)
If the sampling assembly CSA420 is used, 10, 30
or
50 ml are the only values that are permitted.
Conductive sensor
1)
Options
Low sensitivity
Medium sensitivity
High sensitivity
Factory setting
Medium sensitivity
The switching behavior can be set with the liquid
detection function. Use the high sensitivity setting
if the sample has a low conductivity, for example.
Dosing chamber
1)
Options
Dose without pressure (A)
Dose with pressure (B)
Factory setting
Dose without pressure (A)
Dosing with pressure e.g. in conditions with low
suction heights and slight counterpressure or low
volumes.
Path: Menu/Setup/General settings