Operation Liquistation CSF48
84 Endress+Hauser
Liquid detection
2)
Options
•Automatic
• Semi automatic
•Off
Factory setting
Automatic
If "Semiautomatic" is selected, the purge times and
intake times can be defined separately.
Off:
The definition of the purge times and intake times
is completely time-controlled.
Automatic:
The last intake time determined is the new purge
time.
Semi automatic:
If the suction heights tend to vary greatly.
Rinse cycles
2)
0 to 3
Factory setting
0
The suction line is rinsed with the sample up to 3
times.
Safety interlock
2)
(optional) Options
•Off
•On
Factory setting
Off
If the peristaltic pump is opened, the safety
interlock stops all the functions.
Samping time
4)
5 to 20 s
Factory setting
5 s
Enter the time during which the injected piston is
located in the sample.
Dosing time
4)
5 to 100 s
Factory setting
5 s
Enter the time until the sample is dosed.
Dosing mode
4)
Options
• Dose without pressure (A)
• Dose with pressure (B)
Factory setting
Dose without pressure (A)
Dose with pressure (B):
The sample is conveyed with pressure from the
sample compartment of the assembly into the
sampler.
This function is only possible if the compressed air
valve is connected to binary output 1.
Time after sampling
4)
(is only
displayed if more than one
bottle is present)
1 to 100 s
Factory setting
5 s
The time the tap waits in the bottle position when
dosing has finished
Sample pre shots
4)
(is only
displayed if more than one
bottle is present)
0 to 10
Factory setting
0
Enter the sample rinsing cycles with the current
sample.
Assembly cleaning
4)
(is only
displayed if more than one
bottle is present)
Options
•Off
•with air
•with water
Factory setting
Off
Select the purge medium: air connected to binary
output 1 or water connected to binary output 2. If
"Off" is selected, the binary outputs can be used as in
the standard version.
Path: Menu/Setup/General settings
Function Options Info