Liquistation CSF48 Operation
Endress+Hauser 89
PROFIBUS DP
Modbus
Path: Menu/Setup/General settings/Extended setup/PROFIBUS
Function Options Info
Enable Options
•Off
•On
Factory setting
Off
You can switch off PROFIBUS communication at this
point. The software can then only be accessed via local
operation.
Termination Read only If the device is the last in the bus, you can terminate via
the hardware.
--> "Electrical connection" section
Bus address 1 to 125 If you have addressed the bus via hardware (DIP
switches on the module, --> "Connecting digital
communication" section), you can only read the address
here.
If an invalid address is set via the hardware, you have to
assign a valid address for your device either here or via
the bus.
Ident number Options
•Automatic
• PA-Profile 3.02 (9760)
• Liquiline CM44x (155D)
• Liquiline CM44xR (155D)
• Liquistation CSFxx
(155C)
Factory setting
Automatic
Path: Menu/Setup/General settings/Extended setup/Modbus
Function Options Info
Enable Options
•Off
•On
Factory setting
Off
You can switch off Modbus communication at this point.
The software can then only be accessed via local
operation.
Termination Read only If the device is the last in the bus, you can terminate via
the hardware.
--> "Electrical connection" section
Settings
Transmission Mode Options
•TCP
•RTU
•ASCII
The transmission mode is displayed depending on the
version ordered. In the case of RS485 transmission, you
can choose between "RTU" and "ASCII".