Liquisys M CCM223/253 Commissioning
Endress+Hauser 53
Coding Field Range of
adjustment
(factory settings
in bold)
Display Info
O311 Select current
range
4 to 20mA
0 to 20 mA
O311
Sel.Range
4-20
A0025030-EN
O312 0/4 mA value:
Enter the
associated
measured value
0 ppb
0 to 20000 ppb
0.00 ppm
0 to 20 ppm
0.00 mg/l
0 to 20 (5) mg/l
pH 4.00
pH 4 to 9
0 mV
0 to 1500 mV
0 °C
0 to 50 °C
O312
mg/l
0/4 mA
0.00
A0001944-EN
Here you can enter the
measured value at which the
min. current value (0/4 mA) is
applied at the transmitter
output (not for controller).
For the minimum distance
between the 0/4 mA and 20
mA value, see field O313
This field is not displayed if O2
= Contr.
O313 20 mA value:
Enter the
associated
measured value
2000 ppb
0 to 20000 ppb
2.00 ppm
0 to 20 ppm
2.00 (0.50) mg/l
0 to 20 (5) mg/l
pH 9.00
pH 4 to 9
1000 mV
0 to 1500 mV
50 °C
0 to 50 °C
O313
mg/l
20 mA
2.00
A0001946-EN
Here you can enter the
measured value at which the
max. current value (20 mA) is
applied at the transmitter
output.
The minimum distance
between the 0/4 mA and 20
mA value must be as follows:
140/240: 0.2 mg/l
141/241/963: 0.05 mg/l
120: 0.1 mg/l
pH: pH 0.5
ORP: 100 mV
Temperature: 5 °C
This field is not displayed if O2
= Contr.
O3 (2) Simulate
current output
Lin = linear (1)
Sim = simulation
(2)
Tab = table (3)
O3
Sel.Type
sim
A0025039-EN
Simulation is not ended until
O3(1) or O3(3) is selected.
For further characteristics, see
O3 (1), O3 (3).
O321 Enter
simulation
value
Current value
0.00 to 22.00 mA
O321
mA
Simulat.
10.20
A0025040-EN
Entering a current value
results in this value being
directly output at the current
output.
O3 (3) Enter current
output table
Lin = linear (1)
Sim = simulation
(2)
Tab = table (3)
O3
Sel.Type
table
A0025041-EN
Only for ES and EP versions.
Values can also be
subsequently added or
modified. The values entered
are automatically sorted by
increasing current value.
For further characteristics, see
O3 (1), O3 (2).