Commissioning Liquisys M CUM223/253
36 Endress+Hauser
B9
Enter gas bubble
barrier
0.1 %
0.1 to 100 °C
Compensates gas bubble formation, which
may arise from small amounts of dissolved
gas in the medium.
0.1 % = no formation of gas bubbles.
100 % = strong gas bubble formation.
For clear media (measured value below
1000 NTU) always set the gas bubble
barrier to 100 %.