Liquisys M CUM223/253 Commissioning
Endress+Hauser 43
O321
Enter simulation
value
Current value
0.00 to 22.00 mA
Entering a current value results in this
value being directly output at the current
output.
O3 (3)
Enter current output
table (only for Plus
Package)
lin = linear (1)
sim = simulation (2)
tab = table (3)
Versions TB and TS only.
Values can also be added or altered at a
later stage. The values entered are
automatically sorted by increasing
current value.
For further characteristics, see O3 (1),
O3 (2).
O331 Select table options
read
edit
O332
Enter number of
table value pairs
1
1to10
Enter the number of pairs from the x and
y value (measured value and current
value) here.
O333
Select table value
pair
1
1 to No. elem.
Assign
The function chain O333 to O335 will run
through as many times as correspond to
the value in O332. "Assign" appears as the
last step. After confirmation the system
jumps to O336.
O334 Enter x value
0.000 FNU
0.000 NTU
0.000 ppm
0.000 mg/l
0.000 g/l
0.000 kg/l
0.000 t/m
0.000 %
0.000 °C
x value = measured value specified by
user.
O335 Enter y value
4.00 mA
0.00 to 20.00 mA
y value = current value belonging to O334
specified by user.
Return to O333 until all values are
entered.
O336
Message as to
whether table status
is OK
yes
no
Back to O3.
If status = no, correct table (all settings
made up until now are retained) or back
to measuring mode (table is deleted).
Coding Field Setting range
(Factory settings, bold)
Display Info