6.4.7 Concentration measurement
The basic version does not include functions in italic.
R259
Enter number of
cleaning cycles
without cleaning
agent (economy
function)
0
0to9
After cleaning with cleaner, up to 9
cleaning sessions can be carried out with
water only until the next cleaning session
with cleaner takes place.
Coding Field Setting range
(Factory settings, bold)
Display Info
Coding Field Selection or range
(factory settings bold)
Display Info
K
Function group
CONCENTRATION
Four different concentration curves can be
entered in this function group.
K1
Selection of
concentration curve,
to be used to
calculate the display
value
1
1 to 4
The curves are independent of each other.
Therefore, four different curves can be
defined.
K2
Selection of table to
be edited
2
1 to 4
The modification of a table (curve) is
independent from the selected curve in K1.
The selected table (curve) will not be
overwritten until the input of all table
values is complete.
K3 Select table option
read
edit
This selection applies to the concentration
curve selected in K2.
K4
Enter number of
reference elements
1
1 to 10
Each element consists of two numeric
values.
K5 Select element
1
1 to number of element in K4
assign
Any element can be edited.
K6 Enter turbidity value
0.00 %
entire measuring range
The function chain K5 to K7 will be
repeated automatically as many times as
corresponds to the value in K4. Then the
system jumps to K8.