Liquisys M CUM223/253 Commissioning
Endress+Hauser 61
6.4.8 Service
Coding Field Setting range
(Factory settings, bold)
Display Info
S
SERVICE function
group
Service function settings.
S1 Select language
ENG = English
GER = German
FRA = French
ITA = Italian
NL = Dutch
ESP = Spanish
S2 Configure Hold
S+C = Hold during
configuration and calibration
Cal = Hold during calibration
Setup = Hold during
configuration
None = no Hold
S = setup
C = calibration
S3 Manual Hold
off
on
The setting is retained even in the event of
a power failure.
S4
Enter Hold dwell
period
10 s
0 to 999 s
S5
Enter SW upgrade
release code (Plus
Package)
0
0000 to 9999
You can find the code on the nameplate.
If an incorrect code is entered, you are
taken back to the measurement menu.
The number is edited with the PLUS or
MINUS key and confirmed with the ENTER
key.
"1" is displayed if the code is active.
S6
Enter SW upgrade
release code
Chemoclean
0
0000 to 9999
You can find the code on the nameplate.
If an incorrect code is entered, you are
taken back to the measurement menu.
The number is edited with the PLUS or
MINUS key and confirmed with the ENTER
key.
"1" is displayed if the code is active.
S7
Order number is
displayed
CUM253-T
If the device is upgraded, the order code is
not automatically adjusted.
S8
Serial number is
displayed
880CB405G
S
SERVICE
S1
Language
ENG
S2
Auto HOLD
S+C
S3
Man. HOLD
off
S4
s
Cont.Time
10
S5
0
PlusCode
S6
0
CleanCode
S7
order
CUM253-T
S8
SerNo
880CB405G