Commissioning Liquisys M CUM223/253
62 Endress+Hauser
6.4.9 E+H Service
S9
Reset the device to
the basic settings
no
Facty = factory settings
Facty = all data (apart from A1 a. S1) are
deleted and reset to the factory setting!
S10 Perform device test
no
Displ = display test
Coding Field Setting range
(Factory settings, bold)
Display Info
S9
S.Default
no
S10
Test
no
Coding Field Setting range
(Factory settings, bold)
Display Note
E
E+H SERVICE
function group
Information on the device version
E1 Select module
Contr = controller (1)
Trans = transmitter (2)
Main = power unit (3)
Rel = relay module (4)
E111
E121
E131
E141
Software version is
displayed
If E1 = contr: instrument software
If E1 = trans, main, rel: module firmware
E112
E122
E132
E142
Hardware version is
displayed
Only display function
E113
E123
E133
E143
Serial number is
displayed
Only display function
E114
E124
E134
E144
Module ID is
displayed
Only display function
E
E+H SERV
E1
Select
Contr
E111
SW-Vers.
xx.xx
E112
HW-Vers.
xx.xx
E113
SerNo
00044164M
E114
LSC2
Modul-ID