Commissioning Liquisys M CUM223/253
70 Endress+Hauser
C132
Enter concentration
of the second
calibration solution
Current value from C112
C132 1.1 x C131
C133
Enter concentration
of the third
calibration solution
Current value from C113
C133 1.1 x C132
C134
Calibration status is
displayed
o.k.
Exxx
C135
Store calibration
results?
yes
no
new
If C134 = E xxx, then only no or new
(Exception: calibration warning E84). If
new, return to C. If yes / no, return to
“Measurement”.
C1 (4) Select calibration
3-Pt = Three-point
calibration (1)
Corr = Three-point
correction (2)
Edit = Edit calibration (3)
Refl = Fitting with
reflection compensation
(4)
1-Pt = Single-point
calibration (5)
Data = Calibration data (6)
Only for solutions = 2 FNU / 5 ppm!
Backscatter from the immediate sensor
environment is compensated for clear
media.
C141
Enter correct
measured value
0.0 NTU
0.0 to 2.0 NTU
0.0 FNU
0.0 to 2.0 FNU
0.0 ppm
0.0 to 5.0 ppm
0.0 mg/l
0.0 to 5.0 mg/l
Only for the ranges NTU, FNU, ppm, mg/l
C142
Calibration status is
displayed
o.k.
Exxx
C143
Store calibration
results?
yes
no
new
If C142 = E xxx, then only no or new
(Exception: calibration warning E84). If
new, return to C. If yes / no, return to
“Measurement”.
Coding Field Selection or range
(factory settings bold)
Display Info