Technical data Liquisys M CUM223/253
98 Endress+Hauser
Controller
Alarm
Protocol specific data
Function (adjustable): pulse length/pulse frequency controller
Controller response: PID
Control gain K
p
: 0.01 to 20.00
Integral action time T
n
: 0.0 to 999.9 min
Derivative action time T
v
: 0.0 to 999.9 min
Period for pulse length controller: 0.5 to 999.9 s
Frequency for pulse frequency controller: 60 to 180 min
-1
Basic load: 0 to 40% of max. set value
Function (selectable): Latching / momentary contact
Alarm threshold adjustment range: Turbidity / suspended solids / temperature:
complete measuring range
Alarm delay: 0 to 2000 s
0 to 2000 min
HART
Manufacturer ID 11
h
Device type code 0095
h
Transmitter specific revision 0001
h
HART specification 5.0
DD files www.products.endress.com/hart
Load HART 250
Device variables None (dynamic variables PV, SV, only)
Features supported -
PROFIBUS PA
Manufacturer ID 11
h
Ident number 1517
h
Device revision 11
h
Profile version 2.0
GSD files www.products.endress.com/profibus
GSD file version
Output values Main value, temperature value
Input values Display value of PLC
Features supported Device locking: The device can be locked by hardware or
software.
PROFIBUS DP
Manufacturer ID 11
h
Ident number 151F
h
Profile version 2.0
GSD files www.products.endress.com/profibus
GSD file version
Output values Main value, temperature value
Input values Display value of PLC
Features supported Device locking: The device can be locked by hardware or
software.