Diagnostics and troubleshooting Micropilot FMR51, FMR52 FOUNDATION Fieldbus
102 Endress+Hauser
Diagnostic
number
Short text Remedy instructions Status
signal
[from the
factory]
Diagnostic
behavior
[from the
factory]
485 Simulation measured
value
Deactivate simulation C Warning
494 Switch output
simulation
Deactivate simulation switch
output
C Warning
495 Simulation diagnostic
event
Deactivate simulation C Warning
497 Simulation block output Deactivate simulation C Warning
585 Simulation distance Deactivate simulation C Warning
586 Record map Recording of mapping
please wait
C Warning
Diagnostic of process
801 Energy too low Increase supply voltage S Warning
825 Operating temperature 1. Check ambient temperature
2. Check process temperature
F Alarm
921 Change of reference 1. Check reference configuration
2. Check pressure
3. Check sensor
S Warning
941 Echo lost Check parameter 'DC value' F Alarm
1)
942 In safety distance 1. Check level
2. Check safety distance
3. Reset self holding
S Alarm
1)
943 In blocking distance Reduced accuracy
Check level
S Warning
950 Advanced diagnostic 1
to 2 occured
Maintain your diagnostic event M Warning
1)
1) Diagnostic behavior can be changed.
13.7 Event logbook
13.7.1 Event history
A chronological overview of the event messages that have occurred is provided in the
Event list submenu
6)
.
Navigation path
Diagnostics → Event logbook → Event list
A maximum of 100 event messages can be displayed in chronological order.
The event history includes entries for:
• Diagnostic events
• Information events
6) This submenu is only available for operation via local display. In the case of operation via FieldCare, the event list can be displayed with the "Event
List / HistoROM" functionality of FieldCare.