Micropilot FMR56, FMR57 FOUNDATION Fieldbus Operating menu
Endress+Hauser 149
Maximum value
Navigation  Setup → Advanced setup → Linearization → Maximum value
Prerequisite Linearization type (→  146) has one of the following values:
Linear
Pyramid bottom
Conical bottom
Angled bottom
Horizontal cylinder
Sphere
Description Specify the maximum content of the vessel (100%) measured in the units after
linearization.
User entry –50 000.0 to 50 000.0 %
Factory setting 100.0 %
Diameter
Navigation  Setup → Advanced setup → Linearization → Diameter
Prerequisite Linearization type (→  146) has one of the following values:
Horizontal cylinder
Sphere
Description Specify tank diameter.
User entry 0 to 9 999.999 m
Factory setting 2 m
Additional information The unit is defined in the Distance unit parameter (→  127).
Intermediate height
Navigation  Setup → Advanced setup → Linearization → Intermed. height
Prerequisite Linearization type (→  146) has one of the following values:
Pyramid bottom
Conical bottom
Angled bottom
Description Specify intermediate height H.
User entry 0 to 200 m
Factory setting 0 m