EasyManua.ls Logo

Endress+Hauser nivotester FTC 420 - 100 C; Calibration for Maximum Fail-Safe Mode; Calibration for Minimum Fail-Safe Mode; Maximum Fail-Safe Mode

Endress+Hauser nivotester FTC 420
36 pages
Go to English
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Capacitance range seeking Before proceeding, check that the green LED is lit.
Step Procedure
1 Turn the adjuster anticlockwise to the stop: zero setting.
2 Select Range I at the hook switch (Table 3).
3 Turn the adjuster clockwise through the full range and back:
- If the red LED lights and extinguishes, the correct range
is selected. Now calibrate at minimum or maximum
fail-safe mode.
- If not proceed with Step 4
4 Select Range II at the hook switch.
5 Turn the adjuster clockwise through the full range
and back:
- If the red LED lights and extinguishes, the correct range
is selected. Now calibrate at minimum or maximum
fail-safe mode.
- If not proceed with Step 6
6 Select Range III at the hook switch, now calibrate at
mimimum or maximum failsafe mode.
Procedure:
Capacitance range seeking
Calibration for minimum
fail-safe mode
If there is no bridge between terminals 12 and 13 of the Nivotester.
Step Procedure
1 Turn the adjuster switch back to zero.
2 Slowly turn the switch clockwise until the red LED lights.
3 When the exact switching position has been found, turn the
adjuster a further half-division clockwise.
4 When adjusted, check that the unit is functioning correctly
by filling or emptying the vessel..
- For »Trouble-shooting«, see Page 36.
Procedure:
Calibration for minimum
fail-safe mode with uncovered
probe
Calibration for maximum
fail-safe mode
If there is a bridge between terminals 12 and 13 of the Nivotester.
Step Procedure
1 Turn the adjuster switch back to zero.
2 Slowly turn the switch clockwise until the red LED goes out
3 When the exact switch position has been found, turn the
adjuster a further half-division clockwise.
4 When adjusted, check that the unit is functioning correctly
by filling or emptying the vessel..
- For »Trouble-shooting«, see Page 36.
Procedure:
Calibration for maximum
fail-safe mode with uncovered
probe
Nivotester FTC 420/421 Calibration
32

Table of Contents

Other manuals for Endress+Hauser nivotester FTC 420

Related product manuals