4. Trouble-shooting
Normal operation When the Nivotester FTC 420 or 421 is operating:
• the green LED lights continuously,
• the red LED lights when the relay output is de-energised.
The switching of the relay depends upon the fail-safe mode
selected see Figs 7 and 8 on page 28.
Testing the measuring
system
a) Testing the Nivotester FTC 420/421 alone:
• If the FTC 420/421 signals »empty«, short-circuiting terminals
8 and 10 must lead to a »full« signal.
• If the FTC 420/421 signals »full«, which is only possible when a
probe with electronic insert EC 61 is connected, short-circuiting
terminals 8 and 9 leads to an »empty« signal.
b) Testing the complete measuring system:
• Touching the central fixing screw of the electronic insert with an
insulated (electrician’s) screwdriver: the relay in the Nivotester
must switch.
BA 021F/00/a2/05.98
HD/CV4.2
Fault Possible reason
Green LED
does not light
No power - check wiring
Nivotester does
not switch
Wiring/probe fault
- check as described below
Nivotester
switches too late
New product? New electronic insert?
Calibrate again.
For FTC 421 - check timer
Red LED
remains lit/unlit
Build-up at probe
Bad calibration - repeat
Table 5
Trouble-shooting
Endress+Hauser Sales Centers
A
Tel. (01) 88056-0, Fax (01) 8 8056-35
B+L
Tel. (02) 2480600, Fax (02) 2480553
CAN
Tel. (905) 6819292, Fax (905) 6819444
CH
Tel. (061) 7156222, Fax (061) 7111650
D
Tel. (07621) 97501, Fax (07621) 975555
DK
Tel. (31) 6731 22, Fax (31) 67 3045
ES
Tel. (93) 4803366, Fax (93) 4733839
F
Tel. 89696768, Fax 89694802
GB
Tel. (0161) 2865000, Fax (0161) 9981841
HK
Tel. 25 28 3120, Fax 286541 71
I
Tel. (02) 92106421, Fax (02) 92107153
J
Tel. (0422) 540611, Fax (0422) 5502 75
MAL
Tel. (03) 7334848, Fax (03) 7338800
N
Tel. (0 32) 85 1085, Fax (032) 851112
NL
Tel. (035) 6958611, Fax (035) 6958825
S
Tel. (08) 6261600, Fax (08) 6269477
SF
Tel. (90) 8596155, Fax (90) 8596055
SGP
Tel. 4688222, Fax 4666848
THA
Tel. (2) 9967811-20, Fax (2) 9967810
USA
Tel. (317) 535-7138, Fax (317) 535-1489
ZA
Tel. (011) 4441386, Fax (011) 4441977
INTERNATIONAL
Tel. + Fax: see »D«
http://www.endress.com
10.97
0
15399- 0000
Nivotester FTC 420/421 Trouble-shooting