Electrical connection Flowmeter Proline 200 PROFINET with Ethernet-APL
30 Endress+Hauser
8.
A0034176
 7 Sample graphic
9.
10.
A0034177
 8 Sample graphic
Connecting cable (standard, reinforced)
8. Disconnect the signal cable from the connection board of the wall housing . by pressing
in the locking clip on the connector. Remove the transmitter housing.
9. Guide the connecting cable through the cable entry and into the connection housing (if
using a connecting cable without an M12 device plug, use the shorter stripped end of
the connecting cable).
10. Wire the connecting cable:
Terminal 1 = brown cable
Terminal 2 = white cable
Terminal 3 = yellow cable
Terminal 4 = green cable