Proline Promag 10 Installation
Endress+Hauser 13
2.3 Installing the Promag E sensor
"
Caution!
• The plates mounted on the two sensor flanges protect the PTFE which is turned over the
flanges and, consequently, should not be removed until immediately prior to mounting the
sensor.
• The protective plates must always remain mounted while the device is in storage.
• Make sure that the lining at the flange is not damaged or removed.
2.3.1 Seals
Comply with the following instructions when installing seals:
• No seals are required for PFTE measuring tube lining.
• For DIN flanges, only use seals to DIN EN 1514-1.
• Make sure that the mounted seals do not protrude into the piping cross-section.
"
Caution!
Risk of short circuit! Do not use electrically conductive sealing compounds such as graphite! An
electrically conductive layer could form on the inside of the measuring tube and short-circuit the
measuring signal.
2.3.2 Ground cable
If necessary, special ground cables can be ordered as accessories for potential equalization.
2.3.3 Screw tightening torques (Promag E)
Please note the following:
• The tightening torques listed below are for lubricated threads only.
• Always tighten the screws uniformly and in diagonally opposite sequence.
• Overtightening the screws will deform the sealing faces or damage the seals.
• The tightening torques listed below apply only to pipes not subjected to tensile stress.
a0008165
Note!
Screws, nuts, seals, etc. are not included in the
scope of supply and must be supplied by the
customer.
The sensor is installed between the two pipe flanges:
• The requisite torques must be observed ä 14
• If grounding disks are used, follow the mounting
instructions which will be enclosed with the
shipment