Proline Promag 10 Installation
Endress+Hauser 33
"
Caution!
• The ambient temperature range (–20 to +60°C) may not be exceeded at the mounting
location. Avoid direct sunlight.
• If the device is mounted to a warm pipe, make sure that the housing temperature does not
exceed +60 °C, which is the maximum permissible temperature.
Mount the transmitter as illustrated in the diagram.
A0010719
Fig. 1: Mounting the transmitter (remote version)
A Mounted directly on the wall
B Pipe mounting
2.9 Post-installation check
• Is the measuring device damaged (visual inspection)?
• Does the device correspond to specifications at the measuring point, including process
temperature and pressure, ambient temperature, minimum fluid conductivity, measuring
range, etc.?
• Does the arrow on the sensor nameplate match the actual direction of flow through the pipe?
• Is the position of the measuring electrode plane correct?
• Is the position of the empty pipe detection electrode correct?
• Were all screws tightened to the specified torques when the sensor was installed?
• Were the correct seals used (type, material, installation)?
• Are the measuring point number and labeling correct (visual inspection)?
• Were the inlet and outlet runs respected?
– Inlet run 5 × DN
–Outlet run 2 × DN
• Is the measuring device protected against moisture and direct sunlight?
• Is the sensor adequately protected against vibration (attachment, support)?
Acceleration up to 2 g by analogy with IEC 600 68-2-8