Wiring Proline Promag 10
36 Endress+Hauser
3.2 Connecting the remote version connecting cable
3.2.1 Connecting cable for Promag E/L/P/W
Connecting cable termination
Terminate the signal and coil current cables as shown in the figure below (Detail A).
Fit the fine-wire cores with cable end ferrules (Detail B).
Signal cable termination
Make sure that the cable end ferrules do not touch the wire shields on the sensor side! Minimum
distance = 1 mm (0.04 in), exception "GND" = green cable.
Coil current cable termination
Insulate one core of the three-core cable at the level of the core reinforcement; you only require
two cores for the connection.
Sensor Transmitter
mm (inch)
a* = 170 (6.69")
b* = 20 (0.79")
c = 80 (3.15")
d = 17 (0.67")
e = 8 (0.31")
f = 50 (1.97")
g = 1 (0.04")
a0008983
mm (inch)
a* = 100 (3.94")
b = 80 (3.15")
c = 50 (1.97")
d = 17 (0.67")
e = 8 (0.31")
m
= Cable end ferrules, red, 1.0 mm (0.04");
n
= Cable end ferrules, white, 0.5 mm (0.02")
* = Stripping for armored cables only
Sensor Transmitter
mm (inch)
a* = 160 (6.3")
b* = 20 (0.79")
c = 70 (2.76")
d = 50 (1.97")
e = 8 (0.31")
f = 10 (0.39")
a0008984
mm (inch)
a* = 90 (3.54")
b = 70 (2.76")
c = 50 (1.97")
d = 10 (0.39")
e = 8 (0.31")
m
= Cable end ferrules, red, 1.0 mm (0.04");
n
= Cable end ferrules, white, 0.5 mm (0.02")
* = Stripping for armored cables only