PROline Promag 10 3 Installation
Endress+Hauser 15
Inlet and outlet runs
If possible, install the sensor well clear of fittings such as valves, T-pieces, elbows, etc.
Compliance with the following inlet and outlet runs is necessary in order to ensure
measuring accuracy.
Inlet run: 5 x DN
Outlet run: 2 x DN
Fig. 11: Inlet and outlet run
3.2.4 Vibrations
Secure the piping and the sensor if vibration is severe.
"
Caution!
It is advisable to install sensor and transmitter separately if vibration is excessively
severe. Information on resistance to vibration and shockcan be found on Page 80.
Fig. 12: Measures to prevent vibration of the measuring device
F06-10xxxxxx-11-00-00-xx-005F06-10xxxxxx-11-00-00-xx-006