PROline Promag 10 3 Installation
Endress+Hauser 33
3.3.7 Mounting the transmitter (remote version)
The transmitter can be mounted in the following ways:
Wall mounting
Pipe mounting (with separate mounting kit, accessoriesPage 64)
The transmitter and the sensor must be mounted separate in the following
circumstances:
Poor accessibility
Lack of space
Extreme fluid/ambient temperatures (temperature ranges: Page 81),
Severe vibration (> 2 g/2 h per day; 10...100 Hz).
"
Caution!
If the device is mounted to warm piping, make certain that the housing temperature
does not exceed the max. permissible value of +60 °C.
Mount the transmitter as illustrated in Fig. 23.
Fig. 23: Mounting the transmitter (remote version)
A = Direct wall mounting
B = Pipe mounting
ANSCHLUSSKLEMMEN - FIELD TERMINALS
248 ±2
238
ANSCHLUSSKLEMMEN - FIELD TERMINALS
Esc
E
-
+
Esc
E
-
+
A
B
F06-10xxxxxx-17-06-xx-xx-002