PROline Promag 10 4 Wiring
Endress+Hauser 37
Cable termination for the remote version
Promag W / Promag P
Terminate the signal and coil current cables as shown in the figure below (Detail A).
Fit the fine-wire cores with cable end sleeves (Detail B).
" Caution!
When fitting the connectors, pay attention to the following points:
• Signal cable → Make sure that the cable end sleeves do not touch the wire shield on the sensor side!
Minimum distance = 1 mm (exception “GND” = green cable).
• Coil current cable → Insulate one core of the three-core wire at the level of the core reinforcement; you only require two cores for the
connection.
TRANSMITTER
Signal cable Coil current cable
SENSOR
Signal cable Coil current cable
A B
80
50
17
8
Esc
E
-
+
grn
F06-10xxxxxx-04-11-08-en-000
BA
70
50
10
8
Esc
E
-
+
F06-10xxxxxx-04-11-08-xx-000
80
50
17
8
≥1
A
B
grn
F06-5xFxxxxx-04-11-08-en-000
A
B
70
50
10
8
F06-5xFxxxxx-04-11-08-xx-000