Installation Proline Promag 50
6 Endress+Hauser
2 Installation
2.1 Transporting to the measuring point
• Transport the measuring device to the measuring point in the original packaging.
• Do not remove the covers or caps until immediately before installation.
2.1.1 Transporting flanged devices DN ≤ 300 (≤ 12")
2.1.2 Transporting flanged devices DN > 300 (> 12")
Use only the metal eyes provided on the flanges to transport, lift or position the sensor in the
piping.
"
Caution!
Do not attempt to lift the sensor with the tines of a fork-lift truck beneath the metal casing!
This would buckle the casing and damage the internal magnetic coils.
A0008153
A0007408
To transport the unit, use slings slung around the process
connections or use lugs (if available).
#
Warning!
Risk of injury! The device can slip.
The center of gravity of the measuring device may be
higher than the holding points of the slings.
Always ensure that the device cannot slip or turn around
its axis.
A0007409
Do not lift measuring devices by the transmitter housing
or the connection housing in the case of the remote
version. Do not use chains as they could damage the
housing.