Proline Promag 50 Installation
Endress+Hauser
7
2.2 Installation conditions
2.2.1 Dimensions
For the dimensions of the measuring device, see the associated Technical Information on the
CD-ROM.
2.2.2 Mounting location
The accumulation of air or formation of gas bubbles in the measuring tube can result in an
increase in measuring errors.
For this reason avoid the following mounting locations in the pipe:
At the highest point of a pipeline. Risk of air accumulating!
Directly upstream from a free pipe outlet in a down pipe.
A
0008154
Installation of pumps
Do not install the sensor on the intake side of a pump. This precaution is to avoid low pressure
and the consequent risk of damage to the lining of the measuring tube. It might be necessary to
use pulse dampers in systems incorporating piston pumps, piston diaphragm pumps or peristaltic
pumps.
Information on the measuring system's pressure tightness and resistance to vibration and shock
can be found in the Operating Instructions of the CD-ROM.
A
0003203
h 2xDN³
Q-Pulse Id
Active 30/09/2014
Page 7 of 52
Vendor Manual - Electromagnetic Flowmeter - Endress and Hauser - Model Proline Promag 50 - Operating Instructions
Q-Pulse Id VM454
Active 30/09/2014
Page 7 of 52