PROline Promag 50 9Trouble-shooting
Endress+Hauser 89
S
$
SENSOR SW DAT
# 032
Sensor:
Error accessing the calibration
values stored in the S-DAT.
1. Check whether the S-DAT is
correctly plugged into the
amplifier board
Page 98,
100
2. Replace the S-DAT if it is
defective.
Spare parts
Page 96.
Before replacing the DAT,
check that the new, replace-
ment DAT is compatible with
the measuring electronics.
Check the:
Spare part set number
Hardware revision code
3. Replace measuring electro-
nics boards if necessary.
Spare parts
Page 96
S
$
A / C COMPATIB.
# 051
The I/O board and the amplifier
board are not compatible.
Use only compatible modules and
boards. Check the compatibility of
the modules used.
Check the:
Spare part set number
Hardware revision code
No. # 1xx
Software error
S
$
GAIN ERROR AMP
# 101
Gain deviation compared to refer-
ence gain is greater than 2%.
Replace amplifier board.
Spare parts
Page 96
S
$
CHECKSUM TOTAL
# 111
Totalizer checksum error 1. Restart the measuring device
2. Replace the amplifier board if
necessary.
Spare parts
Page 96
S
!
A / C COMPATIB.
# 121
Due to different software ver-
sions, I/O board and amplifier
board are only partially compati-
ble (possibly restricted function-
ality).
! Note!
The indication on the display
as notice message appears
only for 30 seconds (with
listing in “Previous system
condition” function).
This condition can occur if only
one electronics board has
been exchanged; the exten-
ded software functionality is
not available. The previously
existing software functionality
is still working and the meas-
urement possible.
Module with lower software ver-
sion has either to be actualizied
by FieldTool with the required soft-
ware version or the module has to
be replaced.
Spare parts
Page 96
No. # 2xx
Error in DAT / no communication
S
$
COMMUNIC. AMP.
# 251
Internal communication error on
the amplifier board.
Replace amplifier board.
Spare parts
Page 96
S
$
COMMUNICATION I/O
# 261
No data reception between
amplifier and I/O board or faulty
internal data transfer.
Check the BUS contacts
Type Error message / No. Cause Remedy / spare part