PROline Promag 50 9Trouble-shooting
Endress+Hauser 89
S
$
SENSOR SW DAT
# 032
Sensor:
Error accessing the calibration
values stored in the S-DAT.
1. Check whether the S-DAT is
correctly plugged into the
amplifier board
→ Page 98,
100
2. Replace the S-DAT if it is
defective.
Spare parts
→ Page 96.
Before replacing the DAT,
check that the new, replace-
ment DAT is compatible with
the measuring electronics.
Check the:
– Spare part set number
– Hardware revision code
3. Replace measuring electro-
nics boards if necessary.
Spare parts
→ Page 96
S
$
A / C COMPATIB.
# 051
The I/O board and the amplifier
board are not compatible.
Use only compatible modules and
boards. Check the compatibility of
the modules used.
Check the:
– Spare part set number
– Hardware revision code
No. # 1xx
→ Software error
S
$
GAIN ERROR AMP
# 101
Gain deviation compared to refer-
ence gain is greater than 2%.
Replace amplifier board.
Spare parts
→ Page 96
S
$
CHECKSUM TOTAL
# 111
Totalizer checksum error 1. Restart the measuring device
2. Replace the amplifier board if
necessary.
Spare parts
→ Page 96
S
!
A / C COMPATIB.
# 121
Due to different software ver-
sions, I/O board and amplifier
board are only partially compati-
ble (possibly restricted function-
ality).
! Note!
– The indication on the display
as notice message appears
only for 30 seconds (with
listing in “Previous system
condition” function).
– This condition can occur if only
one electronics board has
been exchanged; the exten-
ded software functionality is
not available. The previously
existing software functionality
is still working and the meas-
urement possible.
Module with lower software ver-
sion has either to be actualizied
by FieldTool with the required soft-
ware version or the module has to
be replaced.
Spare parts
→ Page 96
No. # 2xx
→ Error in DAT / no communication
S
$
COMMUNIC. AMP.
# 251
Internal communication error on
the amplifier board.
Replace amplifier board.
Spare parts
→ Page 96
S
$
COMMUNICATION I/O
# 261
No data reception between
amplifier and I/O board or faulty
internal data transfer.
Check the BUS contacts
Type Error message / No. Cause Remedy / spare part