Proline Promag 50H, 53H
4 Endress+Hauser
Input
Measured variable Flow velocity (proportional to induced voltage)
Measuring ranges Measuring ranges for liquids
Typically v = 0.01 to 10 m/s (0.03 to 33 ft/s) with the specified accuracy
Operable flow range Over 1000 : 1
Input signal Status input (auxiliary input)
U = 3 to 30 V DC, R
i
= 5 kΩ, galvanically isolated
Configurable for: totalizer(s) reset, measured value suppression, error-message reset
Status input (auxiliary input) with PROFIBUS DP and MODBUS RS485
U = 3 to 30 V DC, R
i
= 3 kΩ, galvanically isolated
Switching level: 3 to 30 V DC, independent of polarity
Configurable for: totalizer(s) reset (MODBUS RS485 only), measured value suppression, error-message reset,
batching start/stop (optional), batch totalizer reset (optional)
Current input (only Promag 53)
active/passive selectable, galvanically isolated, full scale value selectable, resolution: 3 μA,
temperature coefficient: typ. 0.005% o.r./°C (o.r. = of reading)
active: 4 to 20 mA, R
i
150 Ω, max. 24 V DC, short-circuit-proof
passive: 0/4 to 20 mA, R
i
< 150 Ω, max. 30 V DC
Output
Output signal Promag 50
Current output
active/passive selectable, galvanically isolated, time constant selectable (0.01 to 100 s),
full scale value selectable, temperature coefficient: typ. 0.005% o.r./°C (o.r. = of reading), resolution: 0.5 μA
active: 0/4 to 20 mA, R
L
< 700 Ω (HART: R
L
250 Ω)
passive: 4 to 20 mA, operating voltage V
S
: 18 to 30 V DC, R
i
150 Ω
Pulse/frequency output
passive, open collector, 30 V DC, 250 mA, galvanically isolated
Frequency output: full scale frequency 2 to 1000 Hz (f
max
= 1250 Hz), on/off ratio 1:1, pulse width max. 10s
Pulse output: pulse value and pulse polarity selectable, max. pulse width configurable (0.5 to 2000 ms)
PROFIBUS DP interface
Transmission technology (Physical Layer): RS485 in accordance with ANSI/TIA/EIA-485-A: 1998,
galvanically isolated
Profil version 3.0
Data transmission rate: 9,6 kBaud to 12 MBaud
Automatic data transmission rate recognition
Function blocks: 1 × analog Input, 1 × totalizer
Output data: volume flow, totalizer
Input data: positive zero return (ON/OFF), totalizer control, value for local display
Cyclic data transmission compatible with previous model Promag 33
Bus address adjustable via miniature switches or local display (optional) at the measuring device
PROFIBUS PA interface
Transmission technology (Physical Layer): IEC 61158-2 (MBP), galvanically isolated
Profil version 3.0
Current consumption: 11 mA
Permissible supply voltage: 9 to 32 V
Bus connection with integrated reverse polarity protection
Error current FDE (Fault Disconnection Electronic): 0 mA
Function blocks: 1 × analog input, 2 × totalizer
Output data: volume flow, totalizer
Input data: positive zero return (ON/OFF), control totalizer, value for local display
Cyclic data transmission compatible with previous model Promag 33
Bus address adjustable via miniature switches or local display (optional) at the measuring device