Proline Promag 50P, 53P
Endress+Hauser 13
Performance characteristics
Reference operating
conditions
To DIN EN 29104 and VDI/VDE 2641:
Medium temperature: +28°C ± 2 K
Ambient temperature: +22°C ± 2 K
Warm-up period: 30 minutes
Installation:
Inlet run >10 x Dia. (DN)
Outlet run > 5 x Dia. (DN)
Sensor and transmitter grounded.
Sensor centered relative to the pipe.
Maximum measured error Promag 50:
Pulse output: ± 0.5% o.r. ± 1 mm/s (o.r. = of reading)
Current output: plus typically ± 5 µA
Promag 53:
Pulse output: ± 0.2% o.r. ± 2 mm/s (o.r. = of reading)
Current output: plus typically ± 5 µA
Supply voltage fluctuations have no effect within the specified range.
F06-5xxxxxxx-05-xx-xx-xx-000
Max. measured error in % of reading
Repeatability max. ± 0.1% o.r. ± 0.5 mm/s (o.r. = of reading)
Operating conditions
Installation conditions
Installation instructions Mounting location
Correct measuring is possible only if the pipe is full. Avoid the following locations:
Highest point of a pipeline. Risk of air accumulating.
Directly upstream of a free pipe outlet in a vertical pipe.