Proline Promag 50P, 53P
Endress+Hauser 17
Vibrations
Secure the piping and the sensor if vibration is severe.
Caution!
It is advisable to install sensor and transmitter separately if vibration is excessively severe.
Information on resistance to vibration and shock can be found on Page 20.
F06-5xxxxxxx-11-00-00-xx-006
Foundations, supports
If the nominal diameter is 14" (DN 350), mount the transmitter on a foundation of adequate load-
bearing strength.
Caution!
Do not allow the casing to take the weight of the sensor. This would buckle the casing and
damage the internal magnetic coils.
F06-5xFxxxxx-11-05-xx-xx-000
>30ft(10m)