2 Identification PROline Promag 53
10 Endress+Hauser
2.1.2 Nameplate of the sensor
Fig. 2: Nameplate specifications for the “Promag” sensor (example)
1 Ordering code/serial number: See the specifications on the order confirmation for the meanings of the
individual letters and digits.
2 Calibration factor: 0.5328; zero point:
−5
3 Nominal diameter: 4" (DN 100)
Pressure rating: DIN PN 230 psi (16 bar)
4 TMmax 176 °F (+80 °C) (max. fluid temperature)
5 Materials:
– Lining: hard rubber (HG)
– Measuring electrodes: stainless steel 1.4435
6 Additional information (examples):
– EPD/MSU: with Empty Pipe Detection electrode
– R/B: with reference electrode
– EME/AWE: with exchangeable measuring electrodes
– 0.2% CAL: with 0.2% calibration
7 Reserved for information on special products
8 Ambient temperature range
9 Degree of protection
10 Reserved for additional information on device version (approvals, certificates)
11 Flow direction
2.2 CE mark, declaration of conformity
The devices are designed to meet state-of-the-art safety requirements, have been
tested, and left the factory in a condition in which they are safe to operate. The devices
comply with the applicable standards and regulations in accordance with EN 61010
“Protection Measures for Electrical Equipment for Measurement, Control, Regulation
and Laboratory Procedures”.
The measuring system described in this Operating Manual is therefore in conformity
with the statutory requirements of the EC Directives. Endress+Hauser confirms suc-
cessful testing of the device by affixing to it the CE mark.
i
1
2
3
4
5
6
7
9
10
11
8
ABCDEFGHJKLMNPQRSTTAG No.:
Pat. UK EP 219 725 EP 521169
Pat. UK EP 541 878 EP 618 680
IP67/NEMA/Type4X
HG / 1.4435
0.5328 / –5
TMmax.: 80°C
EPD/MSU R/B EME/AWE 0.2% CAL
DN100 DIN PN 16
K-factor:
Materials:
ENDRESS+HAUSER
12345678901
XXXXX-XXXXXXXXXXXX
PROMAG
Order Code:
Ser.No.:
X
–20°C (–4°F) < Tamb < +60°C+140°F
F06-xxxxxxxx-18-05-xx-xx-000