3 Installation PROline Promag 53
14 Endress+Hauser
Transporting flanged devices (D ≥ 14" / DN ≥ 350):
Use only the metal eyes on the flanges for transporting the device, lifting it and position-
ing the sensor in the piping.
"
Caution:
Do not attempt to lift the sensor with the tines of a fork-lift truck beneath the metal casing.
This would buckle the casing and damage the internal magnetic coils.
Fig. 4: Transporting sensors with D ≥ 14" / DN ≥ 350
3.1.3 Storage
Note the following points:
• Pack the measuring device in such a way as to protect it reliably against impact for
storage (and transportation). The original packaging provides optimum protection.
• The permissible storage temperature is 14 to 122 °F / −10 to +50 °C (preferably 68 °F
/ 20 °C).
• Do not remove the protective plates or caps on the process connections until you are
ready to install the device. This is particularly important in the case of sensors with
Teflon linings.
F06-5xFxxxxx-22-xx-xx-xx-001