3 Installation PROline Promag 53
18 Endress+Hauser
Inlet and outlet runs
If possible, install the sensor well clear of fittings such as valves, T-pieces, elbows, etc.
Compliance with the following requirements for the inlet and outlet runs is necessary in
order to ensure measuring accuracy.
Inlet run 5 x D
Outlet run 2 x D
Fig. 11: Inlet and outlet runs
3.2.4 Vibrations
Secure the piping and the sensor if vibration is severe.
"
Caution:
It is advisable to install sensor and transmitter separately if vibration is excessively
severe. Information on resistance to vibration and shock can be found on Page 124.
Fig. 12: Measures to prevent vibration of the measuring device