4 Wiring PROline Promag 53
50 Endress+Hauser
Cable termination for the remote version
Promag H
Terminate the signal and coil current cables as shown in the figure below (Detail A).
Fit the fine-wire cores with cable end sleeves (Detail B).
" Caution!
When fitting the connectors, pay attention to the following points:
Signal cable
Make sure that the cable end sleeves do not touch the wire shield on the sensor side.
Minimum distance = 0.04"/1 mm (exception “GND” = green cable).
Coil current cable
Insulate one core of the three-core wire at the level of the core reinforcement; you only require two cores for the
connection.
On the sensor side, reverse both cable shields approx. 0.59" (15 mm) over the outer jacket. The strain relief ensures an electrical connec-
tion with the connection housing.
TRANSMITTER
Signal cable Coil current cable
SENSOR
Signal cable Coil current cable
F06-5xxxxxxx-04-11-08-ae-000
Dimensions in inches (mm)
F06-5xxxxxxx-04-11-08-xx-000ae
Dimensions in inches (mm)
F06-5xHxxxxx-04-11-08-ae-000
Dimensions in inches (mm)
F06-5xHxxxxx-04-11-08-xx-000ae
Dimensions in inches (mm)