PROline Promag 53 4 Wiring
Endress+Hauser 53
Fig. 32: Connecting the transmitter (stainless-steel field housing). Cable cross-section: max. 16 AWG (2.5
mm
2
)
a Cable for power supply: 85 to 260 V AC, 20 to 55 V AC, 16 to 62 V DC
Term inal No. 1: L1 for AC, L+ for DC
Term inal No. 2: N for AC, L
− for DC
b Signal cable: Terminals Nos. 20–27
→ Page 54
c Ground terminal for protective conductor
d Ground terminal for signal cable shield
e Service adapter for connecting service interface FXA 193 (FieldCheck™, FieldTool™)
f Cover of the connection compartment
Fig. 33: Connecting the transmitter (wall-mount housing). Cable cross-section: max. (2.5 mm
2
)
a Cable for power supply: 85 to 260 V AC, 20 to 55 V AC, 16 to 62 V DC
Term inal No. 1: L1 for AC, L+ for DC
Term inal No. 2: N for AC, L
− for DC
b Signal cable: Terminals Nos. 20–27
→ Page 54
c Ground terminal for protective conductor
d Ground terminal for signal cable shield
e Service adapter for connecting service interface FXA 193 (FieldCheck™, FieldTool™)
f Cover of the connection compartment
b
c
d
a
2
1
L1 (L+)
N (L-)
f
b
a
e
– 27
– 25
– 23
– 21
+ 26
+ 24
+ 22
+ 20
F06-xxxxxxxx-04-06-xx-xx-006F06-xxxxxxxx-04-03-xx-xx-000