PROline Promag 53 6 Commissioning
Endress+Hauser 87
“Pulsating flow” Quick Setup menu
HOME position → F → MEASURED VARIABLE (A)
MEASURED VARIABLE
→ O → QUICK SETUP (B)
QUICK SETUP
→ N → QS PULSATING FLOW (1003)
Function No. Function name Setting to select ( OS )
(Go to next function: press F )
1003 QS PULSATING FLOW YES
After
F is pressed by way of confirma-
tion, the Quick Setup menu calls up all
the subsequent functions in succession.
▼
▼
Basic configuration
2002 DISPLAY DAMPING 1 s
3002 TOTALIZER MODE (DAA) BALANCE (Totalizer 1)
3002 TOTALIZER MODE (DAB) BALANCE (Totalizer 2)
3002 TOTALIZER MODE (DAC) BALANCE (Totalizer 3)
Signal type for “CURRENT OUTPUT 1...n”
4004 MEASURING MODE PULSATING FLOW
4005 TIME CONSTANT 1 s
Signal type for “PULSE/FREQ. OUTPUT 1...n” (for FREQUENCY operating mode)
4206 MEASURING MODE PULSATING FLOW
4208 TIME CONSTANT 0 s
Signal type for “PULSE/FREQ. OUTPUT 1...n” (for PULSE operating mode)
4225 MEASURING MODE PULSATING FLOW
Other settings
8005 ALARM DELAY 0 s
6400 ASSIGN LOW FLOW CUTOFF VOLUME FLOW
6402 ON-VALUE LOW FLOW CUT OFF Recommended setting:
*
)
Full scale values → Page 21
6403 OFF-VALUE LOW FLOW CUTOFF 50%
6404 PRESSURE SHOCK SUPPRESSION 0 s
▼
Back to the HOME position
→ Press and hold down Esc key (X) for longer than three seconds.
→ Repeatedly press and release Esc key (X) → exit the function matrix step by step
On-value
max. full scale (per DN) *
1000
------ -------- -------------- --------------- ----------- -------- -----
≈