6 Commissioning PROline Promag 53
92 Endress+Hauser
6.2.6 Current output: active/pa ssive
The current outputs can be configured as “active” or “passive” by means of various
jumpers on the I/O board or the current submodule.
#
Warning:
Risk of electric shock. Exposed components carry dangerous voltages. Make sure that
the power supply is switched off before you remove the cover of the electronics com-
partment.
1. Switch off power supply.
2. Remove the I/O board → Page 112, 114
3. Set the jumpers in accordance with Fig. 51.
"
Caution:
– Risk of destroying the measuring device. Set the jumpers exactly as shown in
Fig. 51. Incorrectly set jumpers can cause overcurrents that would destroy either
the measuring device or external devices connected to it.
– Note that the position of the current submodule on the I/O board can vary,
depending on the version ordered, and that the terminal assignment in the con-
nection compartment of the transmitter varies accordingly → Page 54.
4. Installation of the I/O board is the reverse of the removal procedure.
Fig. 51: Configuring current outputs with the aid of jumpers (I/O board)
1 Current output 1 with HART
1.1 Active (default)
1.2 Passive
2 Current output 2 (optional, plug-in module)
2.1 Active (default)
2.2 Passive
F06-x3xxxxxx-16-xx-06-xx-001